ORIGINAL ARTICLE
Year : 2015  |  Volume : 15  |  Issue : 3  |  Page : 229-236

A three-dimensional finite element analysis of a passive and friction fit implant abutment interface and the influence of occlusal table dimension on the stress distribution pattern on the implant and surrounding bone


1 Department of Prosthodontics, Yenepoya Dental College, Derlakatte, Mangalore, Karnataka, India
2 Department of Orthodontics, Yenepoya Dental College, Derlakatte, Mangalore, Karnataka, India

Date of Submission12-Mar-2015
Date of Acceptance21-May-2015
Date of Web Publication15-Sep-2015

Correspondence Address:
Hasan Sarfaraz
Department of Prosthodontics, Yenepoya Dental College, Derlakatte, Mangalore, Karnataka
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0972-4052.161559

Rights and Permissions
  Abstract 

Aims: The aim of the study was to evaluate the stress distribution pattern in the implant and the surrounding bone for a passive and a friction fit implant abutment interface and to analyze the influence of occlusal table dimension on the stress generated.
Materials and Methods : CAD models of two different types of implant abutment connections, the passive fit or the slip-fit represented by the Nobel Replace Tri-lobe connection and the friction fit or active fit represented by the Nobel active conical connection were made. The stress distribution pattern was studied at different occlusal dimension. Six models were constructed in PRO-ENGINEER 05 of the two implant abutment connection for three different occlusal dimensions each. The implant and abutment complex was placed in cortical and cancellous bone modeled using a computed tomography scan. This complex was subjected to a force of 100 N in the axial and oblique direction. The amount of stress and the pattern of stress generated were recorded on a color scale using ANSYS 13 software.
Results : The results showed that overall maximum Von Misses stress on the bone is significantly less for friction fit than the passive fit in any loading conditions stresses on the implant were significantly higher for the friction fit than the passive fit. The narrow occlusal table models generated the least amount of stress on the implant abutment interface.
Conclusion : It can thus be concluded that the conical connection distributes more stress to the implant body and dissipates less stress to the surrounding bone. A narrow occlusal table considerably reduces the occlusal overload.

Keywords: Conical connection, friction fit interface, implant abutment interface, occlusal table dimension, passive fit interface, Tri-lobe connection


How to cite this article:
Sarfaraz H, Paulose A, Shenoy K K, Hussain A. A three-dimensional finite element analysis of a passive and friction fit implant abutment interface and the influence of occlusal table dimension on the stress distribution pattern on the implant and surrounding bone. J Indian Prosthodont Soc 2015;15:229-36

How to cite this URL:
Sarfaraz H, Paulose A, Shenoy K K, Hussain A. A three-dimensional finite element analysis of a passive and friction fit implant abutment interface and the influence of occlusal table dimension on the stress distribution pattern on the implant and surrounding bone. J Indian Prosthodont Soc [serial online] 2015 [cited 2019 Jun 15];15:229-36. Available from: http://www.j-ips.org/text.asp?2015/15/3/229/161559


  Introduction Top


Dental implant-supported prostheses have become one of the significant treatment modalities for replacement of teeth, with reported success rates of over 98.2%. [1] The success of dental implants is highly dependent upon the integration between the implant and the intraoral hard/soft tissues. The long term success or survival of dental implants is determined by the transmission of occlusal load and resultant stress distribution in the surrounding bone. Load transfer at the bone-implant interface depends on: (1) The implant geometry and the design of implant abutment connection; (2) the loading protocol and the type of occlusion; (3) the number of implants and position; (4) the quality and quantity of the surrounding bone. [2],[3],[4],[5],[6]

It has been demonstrated that vertical and transverse masticatory loads induce axial forces and bending moments that results in stress gradients in the implant, as well as in the bone. Rieger et al. [7] reported that stresses in the range 1.4-5.0 MPa may be required for healthy maintenance of bone, stresses outside this range have been reported to cause bone resorption. According to Frost's mechanostat concept, [8] bone fractures at 10,000-20,000 microstrains. However, just 20% to 40% of the amount of strain required for fracture (i.e., 4,000 microstrains) may trigger cytokine to activate a resorptive response. A persistent load increases the stress and may provoke micro-fractures and osteoclastic activity in the bone.

There has been a continuous evolution of the implant abutment connection design with the intensions of reducing these stress concentrations. Based on the various prosthetic and biological complications encountered in the clinical scenario and the results of various studies, the initial external hex design which had the interface above the implant and the osseous crest, has evolved into the internal hex with the implant abutment interface being placed more apically and away from the osseous crest. More than 20 designs of the internal connections are currently being marketed. [9] These can be broadly categorized as follows. [10]



The development of these implant abutment interfaces have reduced the amount of stress that is transmitted to the implant or the surrounding bone and therefore considerably reducing the crestal bone loss, although not entirely eliminated indicating a multi-factorial cause. The structural complexity of implant abutment interface design due to the continuous improvement in the geometry has made it difficult to evaluate occlusal forces in the bone around the dental implant and the stresses within the implant. Finite element analysis (FEA) is a useful tool for the prediction of the effects of stress on the implant and its surrounding bone. [11] The use of the finite element method to analyze stress concentrations was initially introduced into implant dentistry by Weinstein et al. in 1976. [12] In FEA, the mechanical performance of the implant abutment interface could be evaluated by Von Misses stresses. Von Misses stress criterion is important to interpret the stresses within the ductile material, such as the implant material; as deformation occurs when the Von Misses stress value exceeds the yield strength. Finite element studies comparing different connections and the effect of the change in occlusal table width dimensions on stress distribution pattern in and around the implant abutment interface are limited. In this study, an attempt is made to compare the stress distribution pattern of a passive fit and friction fit implant abutment interface in different areas of the implant and bone and the influence of occlusal table width on stress distribution.


  Materials and Methods Top


Finite element analysis is a computerized numerical technique used to determine the stress and displacements through a predetermined model. FEA solves a complex problem by dividing it into a series of interrelated simple problems. A mesh is needed in FEA to divide the complex geometry into smaller elements in which the field variables can be interpolated with the use of shape functions. The process of creating the mesh, elements, their respective nodes and defining boundary conditions is referred to as "discretization" of the problem domain. [13]

Construction of geometric model

The study models were constructed using reverse engineering technique in PRO-ENGINEER 05 through three-dimensional (3D) optical scanning and point cloud data extraction. The reverse-engineering process involves measuring an object and then reconstructing it as a 3D model. Two CAD models of implants were constructed with two different types of implant abutment connections currently available in the market, the passive fit or the slip-fit represented by the Nobel Replace Tri-lobe connection (Nobel Replace, Tapered Groovy, RP 4.3 mm × 13 mm) and the friction fit or active fit represented by the Nobel active conical connection (Nobel active, Internal RP 4.3 mm × 13 mm) along with their respective snappy abutments.

The bone structure was modeled through a computed tomography (CT) scan that can provide results closer to a real scenario, because there is a difference in the behavior of stresses in work conducted with elliptical models, cobblestones, and CT scan data. [14] The thickness of the cortical bone was kept 2 mm, and a uniform layer of cortical bone was modeled on the outer surface of the cancellous core. A bone block model was constructed based on a cross-sectional image of the human mandible in the premolar region, 25 mm high, 12 mm wide, and 10 mm thick consisting of a spongy center surrounded by a 2-mm cortical bone.

Three crowns with different occlusal table dimensions were constructed by changing the buccolingual dimensions and keeping mesiodistal and the cervicoocclusal length constant. The dimensions of a mandibular premolar are 7.5 mm buccolingually, 9 mm mesiodistally, and 8 mm cervicoocclusally. The buccolingual dimension with 7.5 mm was considered as ideal, and then crowns with narrow and wider occlusal tables were constructed with 6 mm and 10 mm buccolingual dimension, respectively. They were then placed over the passive connection and the friction connection making a total of six models. The implant abutment complex thus constructed using reverse-engineering technique was then positioned in the cortical and cancellous bone block.

Mesh generation of the model

The 3D models corresponding to the geometric model was meshed using HYPERMESH 10 and then imported into ANSYS 13 software to perform the numerical simulation. All the components were meshed with solid 92 elements. It is a 2 nd order Tetra Element which has a Quadratic displacement behavior and is well suited to model irregular meshes. The element is defined by 10 nodes having 3° of freedom at each node and 3 translations in nodal X, Y, and Z directions. The numbers of nodes in friction connection for ideal, narrow, and wider occlusal table are 80,786, 82,047, and 83,972; and the number elements are 57,082, 57,737, and 58,931, respectively. The numbers of nodes in passive connection for ideal, narrow, and wider occlusal table are 76,330, 77,469, and 79,527; and the number elements are 53,569, 54,137, and 55,414, respectively. Meshed models of passive-fit connection and friction fit connection are shown in [Figure 1]a.
Figure 1: (a) Meshed models of passive fit connection and friction fit connection. (b) The vertical and oblique loading directions on a meshed model of a passive fit connection

Click here to view


Boundary conditions and constraints

In this study, we assumed the implant, abutment, and screws were homogeneous, linear elastic, and isotropic mechanical properties. However, cortical and cancellous bones were treated as anisotropic. Material properties for bone and implant components [Table 1] were collected from reliable resources and published data. [15],[16],[17],[18],[19],[20] The implant was pure titanium, and other components were titanium alloys, with homogeneous and isotropic elastic properties. It was assumed that there is complete osseointegration between the implant and the surrounding bone.
Table 1: Mechanical properties of different materials.

Click here to view


Loading conditions

A distributed force of 100 N was applied onto the top surface of the crown vertically along the long axis and then obliquely at 45° to the longitudinal axis of the implant. For a direct and systematic comparison, the same loading conditions, boundary conditions, and constraints was applied in all the models. The vertical and oblique loading directions on a meshed model of a passive fit connection as an example are shown in [Figure 1]b.


  Results Top


The data obtained from ANSYS calculation can be presented in a stress distribution map with a color scale, which makes it possible to compare directly the stress level in various component structures of all models. The amount of stress and the pattern of stress generated after applying a load of 100 N on each model in vertical and oblique direction were recorded on a color scale. The values obtained are summarized in the [Table 2]. The [Figure 2], [Figure 3] and [Figure 4] shows the Von Misses stresses during vertical and oblique loading with 100 N on a narrow occlusal table in different regions of friction fit and passive fit implant abutment interface and [Figure 5] and [Figure 6] show graphs of Von Misses stress (in MPa) on narrow, ideal, and wider occlusal tables in vertical and oblique loading on implant, implant abutment interface, and bone in friction fit and passive fit connections.
Figure 2: (a) Von Misses stress on an implant with friction fit connection and passive fit connection during vertical loading with a load of 100 N in models with narrow occlusal table. (b) Von Misses stress on an implant with friction fit connection and passive fit connection during oblique loading with a load of 100 N in models with narrow occlusal table

Click here to view
Figure 3: (a) Von Misses stress on an implant abutment interface with friction fit connection and passive fit connection during vertical loading with a load of 100 N in models with narrow occlusal table. (b) Von Misses stress on an implant abutment interface with friction fit connection and passive fit connection during oblique loading with a load of 100 N in models with narrow occlusal table

Click here to view
Figure 4: (a) Von Misses stress on the bone with friction fit connection and passive fit connection during vertical loading with a load of 100 N in models with narrow occlusal table (b) Von Misses stress on the bone with friction fit connection and passive fit connection during oblique loading with a load of 100 N in models with narrow occlusal table

Click here to view
Figure 5: Graph showing Von Misses stress (in MPa) on narrow, ideal, and wider occlusal tables in vertical and oblique loading on implant and implant abutment interface in friction fit and passive fit connections

Click here to view
Figure 6: Graph showing Von Misses stress (in MPa) on narrow, ideal, and wider occlusal tables in vertical and oblique loading on bone in friction fit and passive fit connections

Click here to view
Table 2: Von misses stress (in mpa) on the implant abutment complex with prosthesis having narrow, ideal and wider occlusal tables

Click here to view


From the values given in [Table 2] the following data and results have been obtained:

  • The overall maximum Von Misses stress on the implant is more significant for friction fit than the passive fit implant abutment interface in both vertical and oblique loading for all the models tested
  • At the implant abutment interface and at the neck of the implant, the Von Misses stress was higher for the friction fit. Whereas on the outer surface of the abutment and on the internal surface of the fixture; the passive fit shows lesser stress in both vertical and oblique loading for all the models tested
  • The overall maximum Von Misses stress on the bone is significantly less for friction fit than the passive fit in both vertical and oblique loading for all the models
  • Irrespective of the type of abutment connection used, the maximum Von Misses stress was seen in the region of the cortical or the marginal bone. It is showed that a significant reduction in Von Misses stress was observed at the boundary between cortical bone and cancellous bone in both loading conditions
  • The narrow occlusal table, irrespective of their connection type has reduced the stress generated. This shows that the width of the occlusal table has got a significant influence on the stress generated on the implant, as well as on the bone.



  Discussion Top


The aim of the study was to analyze the influence of two different types of implant abutment connection on the stress distribution pattern in the implant and the surrounding bone. The implant abutment interface that have been analyzed in the study represent two broad categories of implant abutment connection currently available in the market, the passive fit or the slip-fit represented by the Nobel Replace Tri-lobe connection and the friction fit or active fit represented by the Nobel active conical connection. The stress distribution pattern was studied at different occlusal dimension. Six models were constructed in PRO-ENGINEER 05 of the two implant abutment connection for three different occlusal dimensions each. The implant and abutment complex was placed in cortical and cancellous bone modeled using a CT scan. This complex was subjected to a force of 100 N in the axial and oblique direction. The amount of stress and the pattern of stress generated were recorded on a color scale.

The mean values of overall stress on the implant with friction fit connection were 88.725 for vertical load and 258.962 for oblique load whereas on passive fit were 49.805 and 136.249, respectively, indicating that friction fit produced higher overall stress on the implant than the passive fit. The mean values of overall stress on the bone for vertical and oblique loading on friction fit connection were 15.907 and 57.236 whereas on a passive fit connection were 30.680 and 88.983, respectively. This show that the stress generated by passive fit connection on bone is almost double the stress generated by the friction fit connection. The mean values of the overall stresses on the implant and the bone shows that the friction fit connection absorbs more stress and dissipates less stress to the surrounding bone. The larger contact area and deeper position inside the implant for friction fit connection allowed for better stability and broader stress distribution, as has been observed in several other studies. [21],[22],[23]

Conical connections were developed to achieve friction-based fit of the implant components. [24],[25],[26] This frictional fit creates wedging effects to improve the implant abutment joint stability against the lateral force and helps to transfer the loading force along the conical surface to distribute the stress on the implant, ultimately reducing biological and biomechanical complications. [27],[28] The internal conical connections help the abutment screw retain greater preload after repeated loads since the loading stress is not entirely concentrated on the screw as in the external hex butt joint implant systems. The friction-locking mechanics and the solid design of the friction fit connections provided greater resistance to deformation and fracture under oblique compressive loading when compared to the passive fit connection. [29]

In passive fit connection, the cold welding does not occur when the abutments are tightened thus an inevitable gap between the implant and abutment may still exist. [30],[31] This can cause micro-motion at the interface during clinical loading, which in turn may contribute to stress on the screw and therefore loss of preload and loosening of abutment thereby leading to bacterial colonization of the micro gap. The threshold of deleterious micro-motion level asserted by various researchers' lies within the range of 50-150 μm. [32],[33],[34] Beyond these levels of micro-motion, stress concentration may occur around inserted dental implants leading to crestal bone loss.

The mean values of stress on the crestal bone for vertical and oblique loading on friction fit connection were 15.391 and 25.449 whereas on a passive fit connection were 19.024 and 29.674, respectively. The highest stress occurs in the implant's most cervical region when an occlusal load is applied upon an implant, and the load is partially transferred to the bone. This phenomenon is due to one of the principles of engineering, that is, when two materials are in contact with each other, and one of them is loaded, the stresses will be higher at the materials' initial point of contact. This explains why the cervical region of the implant is the site where the greatest micro-deformations occur independently of the type of bone and the design of the implant, the configuration of the prosthesis and the load. [4] The results of the current FEA for the osseointegrated model are in accordance with the findings of Hansson. [35] Using FEA, Hansson showed that a conical implant abutment interface at the level of the bone crest decreases the peak bone-implant interfacial stress as compared with the flat top interface. For the friction fit implant abutment interface, this peak interfacial shear stress was located at some depth in the marginal bone.

The mean values of stress on the apical area of bone for vertical and oblique loading on friction fit connection were 0.019 and 0.021 whereas on a passive fit connection were 0.0215 and 0.0278, respectively. In this study, significantly larger stress values were seen in the neck area versus the apex area among all models in all conditions, which is consistent with the results of other studies. Stresses induced by occlusal load are initially transferred from implant to the cortical bone, and a small amount of remaining stress spreads to cancellous bone. Higher stress values are observed in cortical bone because of higher modulus of elasticity and bone density compared to the cancellous bone.

Richter [36] has reported that the highest stress in the crestal bone is a result of a transverse load and clenching at centric contacts. The width of almost every natural tooth is greater than the width of the implant used to replace the tooth. The greater the width of a transosteal structure, the lesser the magnitude of stress transmitted to the surrounding bone. The cross-sectional shape of the natural tooth at the crest is biomechanically optimized to resist lateral loads, implants, however, are almost round in cross-section, which is less effective in resisting lateral bending loads thereby concentrating loads in the crestal region. [37] The mean values of axial displacement of teeth in the socket are 25-100 μm, whereas the range of motion of osseointegrated dental implants has been reported approximately 3-5 μm. [38] The elastic modulus of the tooth is closest to bone compared to the available implant biomaterials. Hence, under similar loading conditions implant generates greater stresses and strain at the crest of bone than a natural tooth.

From the results of the study, it is shown that the narrow occlusal table, irrespective of their connection type has reduced the stress generated. This shows that the width of the occlusal table has got a significant influence on the stress generated on the implant, as well as on the bone. Typically, a 30% to 40% reduction in the occlusal table in a molar region has been suggested because any dimension larger than the implant diameter can cause cantilever effects and eventual bending moments in single-implant prostheses. [3],[39] A narrow occlusal table reduces the chance of offset loading and increases axial loading, which eventually can decrease the bending moment. Misch has described how a narrow occlusal table can improve oral hygiene and reduce the risk of porcelain fracture. The proposed key factors to control bend overload in posterior restorations were reduced the inclination of cusps, centrally oriented contacts with a 1-1.5 mm flat area, a narrowed occlusal table, and elimination of cantilevers. [40] As the wider occlusal table will increase stress on the abutment screws, the occlusal table should be reduced in width compared with natural teeth in nonesthetic regions of the mouth.

Analysis of finite elements was shown to be a versatile and promising methodology for analyzing stress concentrations in implant dentistry, but it is worth emphasizing that the FEA is an approximate virtual simulation of clinical situations, presenting certain limitations. [41] Hart et al. demonstrated that FEA models with more than 10,420 nodes showed convergent results. [42] The present study models featured an average of 80,021 nodes and 56,145 elements. Therefore, the results derived from this FEA may be considered to be reasonably accurate and acceptable.


  Conclusion Top


Within the limits of the present study, the following conclusions can be derived:

  • The overall maximum Von Misses stress on the implant is significantly more for friction fit connection than the passive fit connection in both vertical and oblique loading in all the models; whereas the overall maximum Von Misses stress on the bone is significantly less for friction fit connection than the passive fit connection in both vertical and oblique loading in all the models. The overall maximum Von Misses stress values on the implant and the bone show that the friction fit connection absorbs more stress and dissipates less stress to the surrounding bone. Further studies on the permissible amount of micro-movement allowed in a passive fit implant abutment interface may need to be conducted
  • Irrespective of the type of abutment connection used the maximum Von Misses stress was seen in the region of the cortical or the crestal bone. It is shown that a significant reduction in Von Misses stress was observed at the boundaries between cortical bone and cancellous bone in both loading conditions because of relatively low elastic modulus of cancellous bone
  • On comparing the stresses on narrow, ideal and wider occlusal tables for both frictions fit connection and passive fit connection, the results show that the narrow occlusal table has the least stress followed by ideal and the wider occlusal tables. The narrow occlusal table irrespective of their connection type has reduced the stress generated. This shows that the width of the occlusal table has got a significant influence on the stress generated on the implant, as well as on the bone.


The friction fit connection is superior to the passive fit connection, as the friction fit creates wedging effects to improve the implant abutment joint stability against the lateral force and helps to transfer the loading force along the conical surface to distribute the stress on the implant, ultimately reducing biological and biomechanical complications. A narrow occlusal table may increase axial loading and decrease nonaxial loading for the implants thereby reducing the stress on the implant, implant abutment interface, and bone. Thus, it is recommended that the size of the occlusal table to be 30% to 40% smaller for molars.

 
  References Top

1.
Strietzel FP, Karmon B, Lorean A, Fischer PP. Implant-prosthetic rehabilitation of the edentulous maxilla and mandible with immediately loaded implants: Preliminary data from a retrospective study, considering time of implantation. Int J Oral Maxillofac Implants 2011;26:139-47.  Back to cited text no. 1
    
2.
Kim Y, Oh TJ, Misch CE, Wang HL. Occlusal considerations in implant therapy: Clinical guidelines with biomechanical rationale. Clin Oral Implants Res 2005;16:26-35.  Back to cited text no. 2
    
3.
Rangert BR, Sullivan RM, Jemt TM. Load factor control for implants in the posterior partially edentulous segment. Int J Oral Maxillofac Implants 1997;12:360-70.  Back to cited text no. 3
    
4.
Vasconcellos LG, Nishioka RS, Vasconcellos LM, Nishioka LN. Effect of axial loads on implant-supported partial fixed prostheses by strain gauge analysis. J Appl Oral Sci 2011;19:610-5.  Back to cited text no. 4
    
5.
Eskitascioglu G, Usumez A, Sevimay M, Soykan E, Unsal E. The influence of occlusal loading location on stresses transferred to implant-supported prostheses and supporting bone: A three-dimensional finite element study. J Prosthet Dent 2004;91:144-50.  Back to cited text no. 5
    
6.
Shelat S, Kularashmi BS, Annapoorani H, Chakravarthy R. Effect of two different abutment types on stress distribution in the bone around an implant under two loading conditions. J Dent Implant 2011;1:80-5.  Back to cited text no. 6
  Medknow Journal  
7.
Rieger MR, Mayberry M, Brose MO. Finite element analysis of six endosseous implants. J Prosthet Dent 1990;63:671-6.  Back to cited text no. 7
    
8.
Hsu ML, Chen FC, Kao HC, Cheng CK. Influence of off-axis loading of an anterior maxillary implant: A 3-dimensional finite element analysis. Int J Oral Maxillofac Implants 2007;22:301-9.  Back to cited text no. 8
    
9.
Binon PP. Implants and components: Entering the new millennium. Int J Oral Maxillofac Implants 2000;15:76-94.  Back to cited text no. 9
    
10.
Prithviraj DR, Muley N, Gupta V. The evolution of external and internal implant-abutment connections: A review. Int Dent Res 2012;2:37-42.  Back to cited text no. 10
    
11.
Jafari K, Vojdani M, Mahdavi F, Heidary H. Finite element analysis of the effect of superstructure materials and loading angle on stress distribution around the implant. J Dent Biomater 2014;1:57-62.  Back to cited text no. 11
    
12.
Weinstein AM, Klawitter JJ, Anand SC, Schuessler R. Stress analysis of porous rooted dental implants. J Dent Res 1976;55:772-7.  Back to cited text no. 12
    
13.
Geng JP, Tan KB, Liu GR. Application of finite element analysis in implant dentistry: A review of the literature. J Prosthet Dent 2001;85:585-98.  Back to cited text no. 13
    
14.
Wakabayashi N, Ona M, Suzuki T, Igarashi Y. Nonlinear finite element analyses: Advances and challenges in dental applications. J Dent 2008;36:463-71.  Back to cited text no. 14
    
15.
Cibirka RM, Razzoog ME, Lang BR, Stohler CS. Determining the force absorption quotient for restorative materials used in implant occlusal surfaces. J Prosthet Dent 1992;67:361-4.  Back to cited text no. 15
    
16.
Borchers L, Reichart P. Three-dimensional stress distribution around a dental implant at different stages of interface development. J Dent Res 1983;62:155-9.  Back to cited text no. 16
    
17.
Richter EJ, Orschall B, Jovanovic SA. Dental implant abutment resembling the two-phase tooth mobility. J Biomech 1990;23:297-306.  Back to cited text no. 17
    
18.
Quaresma SE, Cury PR, Sendyk WR, Sendyk C. A finite element analysis of two different dental implants: Stress distribution in the prosthesis, abutment, implant, and supporting bone. J Oral Implantol 2008;34:1-6.  Back to cited text no. 18
    
19.
Lan TH, Du JK, Pan CY, Lee HE, Chung WH. Biomechanical analysis of alveolar bone stress around implants with different thread designs and pitches in the mandibular molar area. Clin Oral Investig 2012;16:363-9.  Back to cited text no. 19
    
20.
Huang HL, Hsu JT, Fuh LJ, Tu MG, Ko CC, Shen YW. Bone stress and interfacial sliding analysis of implant designs on an immediately loaded maxillary implant: A non-linear finite element study. J Dent 2008;36:409-17.  Back to cited text no. 20
    
21.
Bozkaya D, Muftu S, Muftu A. Evaluation of load transfer characteristics of five different implants in compact bone at different load levels by finite elements analysis. J Prosthet Dent 2004;92:523-30.  Back to cited text no. 21
    
22.
Maeda Y, Satoh T, Sogo M. In vitro differences of stress concentrations for internal and external hex implant-abutment connections: A short communication. J Oral Rehabil 2006;33:75-8.  Back to cited text no. 22
    
23.
Chun HJ, Shin HS, Han CH, Lee SH. Influence of implant abutment type on stress distribution in bone under various loading conditions using finite element analysis. Int J Oral Maxillofac Implants 2006;21:195-202.  Back to cited text no. 23
    
24.
Semper W, Heberer S, Mehrhof J, Schink T, Nelson K. Effects of repeated manual disassembly and reassembly on the positional stability of various implant-abutment complexes: An experimental study. Int J Oral Maxillofac Implants 2010;25:86-94.  Back to cited text no. 24
    
25.
Harder S, Dimaczek B, Açil Y, Terheyden H, Freitag-Wolf S, Kern M. Molecular leakage at implant-abutment connection - in vitro investigation of tightness of internal conical implant-abutment connections against endotoxin penetration. Clin Oral Investig 2010;14:427-32.  Back to cited text no. 25
    
26.
Harder S, Quabius ES, Ossenkop L, Kern M. Assessment of lipopolysaccharide microleakage at conical implant-abutment connections. Clin Oral Investig 2012;16:1377-84.  Back to cited text no. 26
    
27.
Balfour A, O'Brien GR. Comparative study of antirotational single tooth abutments. J Prosthet Dent 1995;73:36-43.  Back to cited text no. 27
    
28.
Norton MR. An in vitro evaluation of the strength of an internal conical interface compared to a butt joint interface in implant design. Clin Oral Implants Res 1997;8:290-8.  Back to cited text no. 28
    
29.
Coppedê AR, Bersani E, de Mattos Mda G, Rodrigues RC, Sartori IA, Ribeiro RF. Fracture resistance of the implant-abutment connection in implants with internal hex and internal conical connections under oblique compressive loading: An in vitro study. Int J Prosthodont 2009;22:283-6.  Back to cited text no. 29
    
30.
Karl M, Taylor TD. Parameters determining micromotion at the implant-abutment interface. Int J Oral Maxillofac Implants 2014;29:1338-47.  Back to cited text no. 30
    
31.
Norton MR. Assessment of cold welding properties of the internal conical interface of two commercially available implant systems. J Prosthet Dent 1999;81:159-66.  Back to cited text no. 31
    
32.
Szmukler-Moncler S, Salama H, Reingewirtz Y, Dubruille JH. Timing of loading and effect of micromotion on bone-dental implant interface: Review of experimental literature. J Biomed Mater Res 1998;43:192-203.  Back to cited text no. 32
    
33.
Brunski JB. Avoid pitfalls of overloading and micromotion of intraosseous implants. Dent Implantol Update 1993;4:77-81.  Back to cited text no. 33
    
34.
Cameron HU, Pilliar RM, MacNab I. The effect of movement on the bonding of porous metal to bone. J Biomed Mater Res 1973;7:301-11.  Back to cited text no. 34
    
35.
Hansson S. A conical implant-abutment interface at the level of the marginal bone improves the distribution of stresses in the supporting bone. An axisymmetric finite element analysis. Clin Oral Implants Res 2003;14:286-93.  Back to cited text no. 35
    
36.
Richter EJ. In vivo horizontal bending moments on implants. Int J Oral Maxillofac Implants 1998;13:232-44.  Back to cited text no. 36
    
37.
Bidez MW, Misch CE. Force transfer in implant dentistry: Basic concepts and principles. J Oral Implantol 1992;18:264-74.  Back to cited text no. 37
    
38.
Schulte W. Implants and the periodontium. Int Dent J 1995;45:16-26.  Back to cited text no. 38
    
39.
Rangert B, Sennerby L, Meredith N, Brunski J. Design, maintenance and biomechanical considerations in implant placement. Dent Update 1997;24:416-20.  Back to cited text no. 39
    
40.
Dhanasekar B, Aparna IN, Neha M, Amit G. Occlusion in implant dentistry-issues and considerations. J Oral Health Community Dent 2012;6:91-6.  Back to cited text no. 40
    
41.
Segundo RM, Oshima HM, da Silva IN, Burnett LH Jr, Mota EG, Silva LL. Stress distribution of an internal connection implant prostheses set: A 3D finite element analysis. Balt Dent Maxillofac J 2009;11:55-9.  Back to cited text no. 41
    
42.
Hart RT, Hennebel VV, Thongpreda N, Van Buskirk WC, Anderson RC. Modeling the biomechanics of the mandible: A three-dimensional finite element study. J Biomech 1992;25:261-86.  Back to cited text no. 42
    


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4], [Figure 5], [Figure 6]
 
 
    Tables

  [Table 1], [Table 2]



 

Top
 
 
  Search
 
Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
Access Statistics
Email Alert *
Add to My List *
* Registration required (free)

 
  In this article
Abstract
Introduction
Materials and Me...
Results
Discussion
Conclusion
References
Article Figures
Article Tables

 Article Access Statistics
    Viewed1514    
    Printed20    
    Emailed0    
    PDF Downloaded281    
    Comments [Add]    

Recommend this journal


[TAG2]
[TAG3]
[TAG4]