CASE REPORT
Year : 2018  |  Volume : 18  |  Issue : 1  |  Page : 76-81

Prosthetic rehabilitation in neurosurgical cranioplasty


Department of Prosthodontics, Government Dental College, Trivandrum, Kerala, India

Date of Submission05-May-2017
Date of Acceptance21-Nov-2017
Date of Web Publication17-Jan-2018

Correspondence Address:
Dr. Thara Maria Joseph
Department of Prosthodontics, Government Dental College, Trivandrum Medical College Campus, Trivandrum, Kerala
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/jips.jips_136_17

Rights and Permissions
  Abstract 


The defects of the skull cause mechanical vulnerability of the brain, esthetic disfigurement, and transmission of vibrations and pulsation of the brain. Subsequent cranioplasty may be required to compensate for the defect and to alleviate various signs and symptoms. When long-term outcome of biomaterial use in pediatric cases is limited, alloplastic cranioplasty in adults are supported by several large case series. This case report narrates cranioplasty using titanium alloplastic implant material.

Keywords: Cranioplasty, prosthesis, reconstruction, titanium implants, trauma


How to cite this article:
Joseph TM, Ravichandran R, Harshakumar K, Lylajam S. Prosthetic rehabilitation in neurosurgical cranioplasty. J Indian Prosthodont Soc 2018;18:76-81

How to cite this URL:
Joseph TM, Ravichandran R, Harshakumar K, Lylajam S. Prosthetic rehabilitation in neurosurgical cranioplasty. J Indian Prosthodont Soc [serial online] 2018 [cited 2019 Nov 18];18:76-81. Available from: http://www.j-ips.org/text.asp?2018/18/1/76/221173




  Introduction Top


Cranial defects may develop from various causes such as congenital and acquired defects such as surgery, trauma, and pathological diseases.[1] Such defects are treated using cranioplasty to shield the underlying brain tissues. It also aids in relieving pain and better appearance leading to improved quality of life. Minor defects of 2–3 cm dimension located immediately above the orbital rim or nasion may require correction entirely for cosmetic reasons. However, defects of more than 8 cm in dimension at the posterior parieto-occipital junction may require repair almost entirely for the purpose of brain protection.[2] The indications for cranioplasty include the social stigma due to disfigurement, mechanical vulnerability, painful postsurgical cranial defects, and those defects which cannot be repaired by other surgical techniques.[3] There are two methods of cranial reconstruction, namely, osteoplastic reconstruction and other using alloplastic implants. The technique adopted by neurosurgeon is the use of decompressive craniectomy (DC) with duraplasty for life-threatening raised intracranial pressure in traumatic brain injury cases. The benefit of DC is reasonably well-documented worldwide. Autogenous bone, alloplastic materials including metals, heat-polymerizing acrylic, polyethylene, and silicone are usually used materials.[2],[3] However, polyether ether ketone (PEEK) is a recently introduced material for cranioplasty which is a proved successful implant material in orthopedics as well. Titanium has been used recently in fabricating cranial prostheses. This metal is strong, but light material which is soft enough to be swaged in a die-counter-die system. Moreover, it can be strain hardened and thus becomes stronger with manipulation. Sheets which are 0.61 mm thick are adequate, and its radiodensity permits most radiographic studies. After the metal prosthesis is shaped, trimmed, and polished, tissue acceptance of the implant is enhanced by anodizing it in a solution of 80% phosphoric acid, 10% sulfuric acid, and 10% water. Hence, this article describes a technique of cranioplasty using titanium implants which is found to be biocompatible, lightweight, and clinically stable. Cranial prosthesis can be made from casts derived by rapid prototyping techniques.


  Case Report Top


A 23-year-old male patient reported to the Department of Prosthodontics, Government Dental College. He was referred from the Department of Neurosurgery, Medical College for rehabilitation of postsurgical defect on the left side of the head caused from firework tragedy [Figure 1], [Figure 2], [Figure 3]. The tragedy resulted in an open head injury with left temporoparietal contusion and a fracture of the squamous part of the temporal bone. Postoperatively, it was found that the patient had a large bony defect on the left side of the skull with no sensorineural dysfunctions. On examination, the defect was found to be approximately 15 cm × 12 cm in size. He was a college-going student and was more concerned about esthetic recontouring of the defect, and it was of utmost challenge. The defect that was to be restored was large, and a computer-aided design and computer-aided manufacturing (CAD/CAM) prosthesis would have been the most probable option, but considering the economic background of the patient, a custom-made titanium cranial prosthesis was planned and was fabricated to recontour the cranial defect.
Figure 1: Preoperative photographs showing left cranial bone defect (frontal view)

Click here to view
Figure 2: Preoperative photographs (lateral view)

Click here to view
Figure 3: Computed tomography with skull defect

Click here to view


For fabrication of prosthesis, outer and inner margins of the defect were located and marked with the help of neurosurgeon before impression making. Irreversible hydrocolloid (Alginate) was mixed vigorously with cold water and spread out over the scalp area. Cotton gauze was partially embedded. After setting, 2–3 layers of quick setting plaster was poured to achieve a firm base [Figure 4], and the impression was removed and poured in dental stone (kalstone) [Figure 5]. As per the instructions of neurosurgeon, markings were made 5 mm beyond the edge of the defect, so that 2-mm diameter holes can be placed to fix the prosthesis with titanium implant.
Figure 4: Impression of the defect

Click here to view
Figure 5: Retrieved cast

Click here to view


The cast was then painted with a suitable separating medium. Molten wax was poured into the defect for the fabrication of a wax pattern, and it was carved out from inside to simulate the thickness of the bone [Figure 6]. The wax over the defect in cast model was carefully contoured and tried for bilateral symmetry. The wax pattern was retrieved and used for the fabrication of titanium mold using a 1.2-mm thick titanium alloy (Ti6Al4V) plate. The titanium plate is not made for full thickness of wax pattern; instead, the outer contour only is simulated. The minimum thickness of the titanium mold is 0.5 mm. The manufacturing process was carried out by applying pressure on titanium plate, which was exerted by two pieces of Duraluminum in which the wax pattern of implant was machined. In the titanium implant, 2-mm holes were drilled with same drill bits throughout the plate as per the instructions of neurosurgeon and were tried extracranially [Figure 7]. Sterilization of implant was done as per the theater sterilization protocol. The patient was under general anesthesia and frequently monitored in the surgical theater. Peroperative, after reflection of flap [Figure 8] and exposure of the bony defect, the implant was made to seat in the defect properly [Figure 9]. The screws were tightened into the holes to secure implant into position, and the flap was sutured back into position [Figure 10]. Sutures were removed on the 7th postoperative day, and the patient was discharged with uneventful healing. The reestablished cranial contour was well appreciated during follow-up [Figure 11] and [Figure 12].
Figure 6: Wax pattern of the defect

Click here to view
Figure 7: Extracranial tryin of titanium plate

Click here to view
Figure 8: Neurosurgical procedure

Click here to view
Figure 9: Neurosurgical procedure showing titanium implant in position

Click here to view
Figure 10: Sutures placed

Click here to view
Figure 11: Postoperative view showing excellent adaptation and symmetry

Click here to view
Figure 12: Postoperative aerial view

Click here to view



  Discussion Top


The physical and mechanical properties of materials used for the fabrication of maxillofacial prosthesis influence success of prosthodontic treatment. The materials commonly used for fabrication of these prosthesis are acrylic resins, acrylic copolymers, various metals, vinyl polymers, polyurethane elastomers, silicone elastomers, and PEEK.[4] However, none of them fulfill all the requirements for a satisfactory prosthesis.[5] The flexibility, good surface texture, and lifelike appearance of silicone make it the most widely used material for fabrication of maxillofacial prosthetic material.[6],[7],[8] Recently, CAD/CAM system for fabrication of oral and maxillofacial prosthesis is being used. However, its complexity, cost, and nonavailability at many centers is a limitation.[7] The advantages of cranioplasty include not only the anatomic accuracy but also the esthetic contouring as well as maintenance of hemodynamics. Most cranioplasty materials are used with little adjustments by the surgeon. However, the basic principle should not be forgotten; to choose proper material for the defect. A cranioplasty material must have low infection rates, show low heat conduction, to be nonmagnetic, radiolucent, tissue acceptable, durable, shapable, and inexpensive.[9] Before achieving bone closure, clear bone borders should be obtained, and scalp should be dissected from dura. Dural tears should be closed in watertight manner. Bone and cranioplasty material should touch to each other with maximum capacity. To prevent the mobility of the cranioplasty flap, the material is fixed to the bone with proper plates. Prefabrication of polymethyl methacrylate (PMMA) prostheses by hand has been used since the 1970s employing various procedures, but these methods appear to have been shadowed by newer CAD/CAM techniques,[10] which basically consist of using imaging from the patients' cranial defect and prefabricating the PMMA prosthesis using a three-dimensional (3D) printer. Certainly, the use of these techniques becomes a challenge for developing and third-world countries, in which limited economic and logistical resources do not allow for the extensive use of such technology. With retrospective studies, titanium has proved to be an excellent material for cranioplasty, especially with the appropriate technology for preparation.[11] Precise communication between the neurosurgeon, maxillofacial surgeon, and prosthetist to give individual consideration of the patient's defect, plate design, and orientation is essential. The custom-made implants for cranioplasty showed a significant improvement in morphology.[12] The implants may be very useful for repairing large- and complex-shaped cranial defect. Using a 1.2-mm thick titanium alloy (Ti6Al4V) plate, the implant was fabricated. The manufacturing process was carried out by applying pressure on titanium plate, which was exerted by two pieces of Duraluminum in which the wax pattern of the implant was machined. In the titanium implant, 2-mm holes were drilled with same drill bits throughout the plate as per the instructions of neurosurgeon. Titanium mold was perforated uniformly, in this case, to prevent fluid accumulation beneath the prosthesis, and enhance lymphatic drainage, to ensure growth of fibrous connective tissue into the holes for better adhesion of scalp over the prosthesis and also to prevent accumulation of epidural hematoma.[13] Titanium is well established in the field of cranial and craniofacial medicine. This is primarily due to its biological inertness, favorable strength-to-weight ratio, and favorable cosmetic and functional outcomes. Another key advantage of titanium is its natural osseointegration factor, which promotes active bone growth into the implant. Despite its lower Young's modulus than alternative metals, the value for titanium is higher than that of natural bone, leading to a lessened, but still present, risk of stress shielding. Titanium implants will often require prefabrication leading to an increase lead time and cost. In addition, intraoperative alteration remains difficult. Studies have shown that titanium implants for cranial and maxillofacial applications have a lower survival rate (higher complication rate)[14] than alternative materials such as PMMA. The manual techniques used are often doubtful about its fit and tissue growth. Autografts, allografts, xenografts, celluloids, methyl methacrylate, hydroxyapatite, silicone, ceramic, metal allografts such as aluminum, gold, silver, palladium, tantalum, and titanium are being used and among this due to various properties of titanium as discussed, they are used for cranioplasty widely. Recent technologies such as 3D computed tomography (CT) and stereolithographic models evaluate the exact volume and contour of the defect and thus ensure exact fit of the prosthesis. However, considering the economic status of the patients, clinicians are often forced to deliver prosthesis using manual methods. Thus, rehabilitation of cranial defect with titanium implants is often a challenge to us. Adequate rehabilitation of skull disfigurement with regained strength on left side of the skull was the highlight of this case report.


  Conclusion Top


Number of patients requiring cranioplasty has increased considerably in recent years due to large number of cranial injuries occurring in this modern age. This case report illustrates a patient, who was a victim of firework tragedy. His defect was corrected by initial surgical correction and cranioplasty with titanium implant. Ongoing researches on both biologic and nonbiologic substitutions continue with the help of recent technology. Stem cell experiments and development of morphogenic proteins are expected to take place in the short-term future. This case has been reviewed after 2 months, and later, follow-up was done again after 8 months [Figure 13] and [Figure 14]. The patient did not report any complications till date and CT scan after 10 months revealed perfect adaptation of mold to the skull without any abnormal findings [Figure 15]. The results were appreciable.
Figure 13: Follow-up 8 months

Click here to view
Figure 14: Follow-up 8 months

Click here to view
Figure 15: Computed tomography scan after 10 months

Click here to view


Declaration of patient consent

The authors certify that they have obtained all appropriate patient consent forms. In the form, the patient has given his consent for his images and other clinical information to be reported in the journal. The patient understand that name and initials will not be published and due efforts will be made to conceal identity, but anonymity cannot be guaranteed.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.



 
  References Top

1.
Pavaiya A, Tyagi VK, Tripathi A, Singh SV, Chand P. Cranioplasty with alloplastic cranial implant. J Indian Prosthodont Soc 2009;9:109-11.  Back to cited text no. 1
  [Full text]  
2.
Ebrahimi A, Nejadsarvari N, Rasouli HR, Ebrahimi A. Warfare-related secondary anterior cranioplasty. Ann Maxillofac Surg 2016;6:58-62.  Back to cited text no. 2
[PUBMED]  [Full text]  
3.
Goiato MC, Pesqueira AA, dos Santos DM, Antenucci RM, Ribeiro Pdo P. Evaluation of dimensional change and detail reproduction in silico nes for facial prostheses. Acta Odontol Latinoam 2008;21:85-8.  Back to cited text no. 3
    
4.
Punchak M, Chung LK, Lagman C, Bui TT, Lazareff J, Rezzadeh K, et al. Outcomes following polyetheretherketone (PEEK) cranioplasty: Systematic review and meta-analysis. J Clin Neurosci 2017;41:30-5.  Back to cited text no. 4
    
5.
Alqutaibi AY. Materials of facial prosthesis: History and advances. Int J Contemp Dent Med Rev 2015;2015:4.  Back to cited text no. 5
    
6.
Mortellaro C, Garagiola U, Lucchina AG, Grivetto F, Milone G, Pappalardo S, et al. The use of silicon elastomer in maxillofacial rehabilitation as a substitute for or in conjunction with resins. J Craniofac Surg 2006;17:152-62.  Back to cited text no. 6
    
7.
Tsuji M, Noguchi N, Ihara K, Yamashita Y, Shikimori M, Goto M, et al. Fabrication of a maxillofacial prosthesis using a computer-aided design and manufacturing system. J Prosthodont 2004;13:179-83.  Back to cited text no. 7
    
8.
Kharade P, Dholam K, Gorakh A. A technique for fabrication of cranial prostheses using high-temperature vulcanizing silicone material. J Prosthet Dent 2017;118:113-5.  Back to cited text no. 8
    
9.
Seckin A, Baris K. Cranioplasty: Review of materials and techniques. J Neurosci Rural Pract 2011;2:162-7.  Back to cited text no. 9
    
10.
Mathew N, Joseph S, Joseph AM, Koshy AT, George R. Rehabilitation of a craniofacial defect using extra-cranial prosthesis. J Int Oral Health 2016;8:813-6.  Back to cited text no. 10
  [Full text]  
11.
Piazza M, Grady MS. Cranioplasty. Neurosurg Clin N Am 2017;28:257-65.  Back to cited text no. 11
    
12.
Goiato MC, Pesqueira AA, Ramos da Silva C, Gennari Filho H, Micheline Dos Santos D. Patient satisfaction with maxillofacial prosthesis. Literature review. J Plast Reconstr Aesthet Surg 2009;62:175-80.  Back to cited text no. 12
    
13.
Parichay PJ, Khanapure K, Joshi KC, Aniruddha TJ, Sandhya M, Hegde AS, et al. Clinical and radiological assessment of cerebral hemodynamics after cranioplasty for decompressive craniectomy – A clinical study. J Clin Neurosci 2017;42:97-101.  Back to cited text no. 13
    
14.
Le C, Guppy KH, Axelrod YV, Hawk MW, Silverthorn J, Inacio MC, et al. Lower complication rates for cranioplasty with peri-operative bundle. Clin Neurol Neurosurg 2014;120:41-4.  Back to cited text no. 14
    


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4], [Figure 5], [Figure 6], [Figure 7], [Figure 8], [Figure 9], [Figure 10], [Figure 11], [Figure 12], [Figure 13], [Figure 14], [Figure 15]



 

Top
 
 
  Search
 
Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
Access Statistics
Email Alert *
Add to My List *
* Registration required (free)

 
  In this article
Abstract
Introduction
Case Report
Discussion
Conclusion
References
Article Figures

 Article Access Statistics
    Viewed1585    
    Printed15    
    Emailed0    
    PDF Downloaded159    
    Comments [Add]    

Recommend this journal


[TAG2]
[TAG3]
[TAG4]