• Users Online: 974
  • Home
  • Print this page
  • Email this page
Home About us Editorial board Ahead of print Current issue Search Archives Submit article Instructions Subscribe Contacts Login 


 
 Table of Contents  
CASE REPORT COMPETITION
Year : 2018  |  Volume : 18  |  Issue : 5  |  Page : 41

CRC14: Different approaches to morse taper abutment fracture


Date of Web Publication31-Oct-2018

Correspondence Address:
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0972-4052.244669

Rights and Permissions

How to cite this article:
. CRC14: Different approaches to morse taper abutment fracture. J Indian Prosthodont Soc 2018;18, Suppl S1:41

How to cite this URL:
. CRC14: Different approaches to morse taper abutment fracture. J Indian Prosthodont Soc [serial online] 2018 [cited 2018 Nov 20];18, Suppl S1:41. Available from: http://www.j-ips.org/text.asp?2018/18/5/41/244669



Ee Lian Lim1, Wei Cheong Ngeow1, Daniel Lim1, Wen Lin Chai2

1Department of Oral and Maxillofacial Clinical Sciences, University of Malaya, 50603 Kuala Lumpur, Malaysia

2Department of Restorative Dentistry, Faculty of Dentistry, University of Malaya, 50603 Kuala Lumpur, Malaysia

Introduction: Conical internal connection (CIC; internal friction type) or the so-called Morse taper design provides a tight junction through the friction between implant fixture and abutment, & not through a screw. However, there are abutments which incorporate both designs to ensure that the abutment would not loosen. Having said so, they are difficult to deal with in the event of abutment fracture.

Case Descriptions: This series of cases highlight fractures associated with Morse taper abutment and its range of managements. The first case was managed by disengaging both the old abutment and its connection screw using a kit manufactured by the supplier, while the second case involved removing the abutment and screw by drilling away the inner chamber of the abutment, followed by the removal of the fractured screw using a modified screw removal kit. The last case involved the removal of the implant concerned by trephining, when both described approaches failed.

Discussion: The addition of a screw to abutments with a Morse taper design makes them very difficult to remove if both are fractured. In the second case, we have to modify a drill stabilising sleeve to engage the screw as there's none commercially available.

Conclusion: This presentation highlights the need of different types of management to Morse taper abutment fracture from merely unscrewing the old abutment (mild approach), to removing the abutment and screw (moderate approach) and finally to removing the implant concerned (severe approach).






 

Top
 
 