ORIGINAL ARTICLE
Year : 2017  |  Volume : 17  |  Issue : 4  |  Page : 388-394

Effect of joining the sectioned implant-supported prosthesis on the peri-implant strain generated in simulated mandibular model


1 KLES's Institute of Dental Sciences, Bengaluru, Karnataka, India
2 Vishnu Dental College, Kovvada, Andhra Pradesh, India
3 AECS Maaruti Dental College, Bengaluru, Karnataka, India

Date of Submission26-May-2017
Date of Acceptance25-Sep-2017
Date of Web Publication3-Nov-2017

Correspondence Address:
Ipsha Singh
#210, 2nd Main, 11th Cross, HIG Colony, RMV 2nd Stage, Bangalore - 560 094, Karnataka
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/jips.jips_144_17

Rights and Permissions
  Abstract 

Aim: The aim of this study is to evaluate the strain developed in simulated mandibular model before and after the joining of an implant-supported screw-retained prosthesis by different joining techniques, namely, arc welding, laser welding, and soldering.
Materials and Methods: A specimen simulating a mandibular edentulous ridge was fabricated in heat-cured acrylic resin. 4-mm holes were drilled in the following tooth positions; 36, 33, 43, 46. Implant analogs were placed in the holes. University of California, Los Angeles, abutment was attached to the implant fixture. Eight strain gauges were attached to the acrylic resin model. Six similar models were made. Implant-supported screw-retained fixed prosthesis was fabricated in nickel-chromium alloy. A load of 400 N was applied on the prosthesis using universal testing machine. Resultant strain was measured in each strain gauge. All the prostheses were sectioned at the area between 36 and 33, 33 and 43, and 43 and 46 using 35 micrometer carborundum disc, and strain was measured in each strain gauge after applying a load of 400 N on the prosthesis. Specimens were joined by arc welding, soldering, and laser welding. After joining, a load of 400 N was applied on each prosthesis and the resultant strain was measured in each strain gauge.
Results: Highest mean strain values were recorded before sectioning of the prostheses (889.9 microstrains). Lowest mean strain values were recorded after sectioning the prosthesis and before reuniting it (225.0 microstrains).
Conclusions: Sectioning and reuniting the long-span implant prosthesis was found to be a significant factor in influencing the peri-implant strain.

Keywords: Arc welding, laser welding, passive fit, peri-implant strain, soldering


How to cite this article:
Singh I, Nair K C, Shetty J. Effect of joining the sectioned implant-supported prosthesis on the peri-implant strain generated in simulated mandibular model. J Indian Prosthodont Soc 2017;17:388-94

How to cite this URL:
Singh I, Nair K C, Shetty J. Effect of joining the sectioned implant-supported prosthesis on the peri-implant strain generated in simulated mandibular model. J Indian Prosthodont Soc [serial online] 2017 [cited 2021 Apr 23];17:388-94. Available from: https://www.j-ips.org/text.asp?2017/17/4/388/217642


  Introduction Top


A precise fit between an implant body and an abutment as well as between an implant abutment and superstructure are important factors in determining the long-term success of an implant-supported prosthesis. Thus, when the fit is not satisfactory, tensile, compressive, and torsional stresses may be introduced into the prosthesis.[1] This may result in loss of osseointegration, loosening of the prosthesis or abutment screws, distortion or breakage of prosthesis, microfractures in the bone surrounding the implant, or fracture of implant body.[1] Watanabe et al. proposed that 90% of the contact surfaces between an abutment and its superstructure will have a gap within the range of 30-micron meter, which is too small to be clinically detectable.[2] Therefore, the cast framework that seems to fit well upon visual examination could undergo distortion or deformation of the contacting surface during the tightening of the prosthesis as well as induce stresses on the bone surrounding the implant.[3] Obtaining passive fit in a long-span prosthesis is a challenge, and various methods have been developed to improve the fit between abutment and its superstructure such as sectioning and joining method (one-piece cast superstructure is cut into pieces corresponding to each abutment and then pieces are reassembled and joined) and use of IMZ “passive-fit system” (utilizes fastening screws, sleeves, plastic sleeves, and titanium coping).[4] The objective of this study was to investigate the strain produced in the bone surrounding implants when the long-span implant-supported prostheses were fabricated using one-piece casting method and to compare it with the strain generated when prostheses were sectioned and reunited by various techniques, namely, soldering, arc welding, and laser welding. The hypothesis of the study was that there is no difference between the peri-implant strain generated on the bone before and after sectioning and joining the long-span implant-supported prosthesis irrespective of the joining technique used.


  Materials and Methods Top


Preparation of the specimen

Specimen consisting of a 10-mm thick flat base and a rim simulating a mandibular edentulous ridge was fabricated in modeling wax [Figure 1]. This was duplicated in heat-cured acrylic resin. 4-mm holes were drilled in the following tooth positions; 36, 33, 43, 46 using tungsten carbide bur (4 mm diameter). Make it simple, implant analogs of dimension 3.75 mm × 11.5 mm were placed in the holes and they were secured in place with autopolymerizing acrylic resin [Figure 2]. University of California, Los Angeles (UCLA) abutment was attached to the implant fixture. All the abutments were fixed to the implant with a torque of 35 Ncm using torque wrench. Implant-supported screw-retained fixed prosthesis framework was fabricated in nickel-chromium alloy connecting all the four implant analogs [Figure 3] and [Figure 4]. All the prostheses were fixed to the abutment with a torque of 35 Ncm using a torque wrench.
Figure 1: Wax for of simulated mandibular model

Click here to view
Figure 2: Heat-cured acrylic model with implants

Click here to view
Figure 3: Screw-retained implant-supported prosthesis

Click here to view
Figure 4: Mandibular model with the prosthesis

Click here to view


Eight strain gauges (resistance 350 ohms, length 3 mm, factor 2.01) were attached to the acrylic model mesial and distal to the implants [Figure 5]. All the strain gauges were set to zero. Strain gauges were numbered according to their position next to the implant. Six similar models were made.
Figure 5: Model with strain gauges

Click here to view


  1. Strain gauge fixed distal to the right first molar implant (46)
  2. Strain gauge fixed mesial to the right first molar implant (46)
  3. Strain gauge fixed distal to the right canine implant (43)
  4. Strain gauge fixed mesial to the right canine implant (43)
  5. Strain gauge fixed mesial to the left canine implant (33)
  6. Strain gauge fixed distal to the left canine implant (33)
  7. Strain gauge fixed mesial to the left first molar implant (36)
  8. Strain gauge fixed distal to the left first molar implant (36).


Grouping specimens

Six specimens were grouped into three - A, B, C with each group consisting of two specimens, according to the joining methods used, i.e., soldering, arc welding, laser welding.

Measurement of the strain

A load of 400 N was applied for a period of 10 s on the prosthesis using universal testing machine. A steel plate of 50 mm diameter was used to ensure uniform contact of the prosthesis during load application. Resultant strain was recorded with respect to each strain gauge.

Sectioning of the prosthesis

All the prosthesis were sectioned between 36 and 33, 33 and 43, and 43 and 46 [Figure 6] using 35 micrometer carborundum disc, and strain was measured in each strain gauge after applying a load of 400 N on the sectioned prosthesis for 10 s.
Figure 6: Sectioned implant-supported prosthesis

Click here to view


Joining of the prosthesis

Sectioned specimens of groups A, B, and C were united by arc welding, soldering, and laser welding, respectively. After joining, a load of 400 N was applied on each prosthesis for a period of 10 s and the resultant strain was recorded in each strain gauge [Figure 7].
Figure 7: Load application on the model using universal testing machine

Click here to view



  Results Top


Results of the present study are given in the following tables. [Table 1] consists of the peri-implant strain values before sectioning of the prostheses. [Table 2] consists of the strain values after sectioning of the prostheses. [Table 3], [Table 4], [Table 5] consists of the strain values after joining of the sectioned prostheses by arc welding, soldering, and laser welding, respectively. [Table 6], [Table 7], [Table 8], [Table 9],[Table 10] contains mean and standard deviation before sectioning, after sectioning, after arc welding, after soldering and after laser welding the prostheses respectively.
Table 1: Peri-implant strain generated in the mandibular model before sectioning of the prostheses (microstrain)

Click here to view
Table 2: Peri-implant strain generated in the mandibular model after sectioning of the prostheses (microstrain)

Click here to view
Table 3: Peri-implant strain generated in the mandibular model after joining the sectioned prostheses by arc welding (microstrain)

Click here to view
Table 4: Peri-implant strain generated in the mandibular model after joining the sectioned prostheses by soldering (microstrain)

Click here to view
Table 5: Peri-implant strain generated in the mandibular model after joining the sectioned prostheses by laser welding (microstrain)

Click here to view
Table 6: Mean and standard deviation before sectioning the prostheses

Click here to view
Table 7: Mean and standard deviation after sectioning the prostheses

Click here to view
Table 8: Mean and standard deviation after arc welding the prostheses

Click here to view
Table 9: Mean and standard deviation after soldering the prostheses

Click here to view
Table 10: Mean and standard deviation after laser welding the prostheses

Click here to view


Statistical analysis

The statistical analysis was performed using the one-way analysis of variance (ANOVA) and Scheffe's post hoc test. The results of the statistical analysis are tabulated from [Table 11].
Table 11: Multiple comparisons (Scheffe's posthoc test)

Click here to view


Test of normality (Shapiro–Wilk) showed a normal distribution (P > 0.05). Hence, parametric test, one-way ANOVA, and Scheffe's post hoc were performed.

The mean difference was significant at the 0.05 level which means that P ≤ 0.05 (significant) is considered as statistically significant.


  Discussion Top


Precise fit between the abutments and superstructure is an important factor in determining the long-term success of an implant-supported prosthesis. Passive fit though desirable is not clinically obtainable. Passive fit means that framework induces zero strain on the implant and the surrounding bone in the absence of an external load.[3] The clinical and laboratory procedures employed in the fabrication of framework are inadequate to provide a passive fitting superstructure. When the passivity in superstructure is not achieved, forces are generated in the bone around the implant which may result in loosening of the prosthesis or abutment screws and fracture of the framework or implant body.[5] Watanabe et al. stated that such a situation may even lead to loss of osseointegration.

[2] According to Vasconcellos et al., when an occlusal load is applied on an implant-supported prostheses, the load is partially transferred to bone, with the highest stress occurring in the peri-implant area.[6] Therefore, the cervical region of implant is the site where the greatest microdeformation occurs independent of the type of bone, the design of implant, the configuration of prosthesis, and the type of load applied. Himmlová et al. stated that bone strain above 3000 microstrains may be unfavorable for the bone leading to a hypertrophic response and bone strain above 4000 microstrains may cause overloading followed by bone loss.[7] Complete passivity in one-piece casting is hard to achieve, but improvement in the fit of implant-supported framework can be achieved by sectioning the framework and then reuniting the sectioned framework.[8] The need for this study was to develop a clinical approach to reduce the stresses induced on the bone surrounding the implant since these stresses on exceeding the physiological limit of the bone can cause crestal bone loss and loss of osseointegration. This study was done to evaluate the strain developed in simulated mandibular model before and after the joining of an implant-supported screw-retained prosthesis by different techniques.

Specimens simulating mandibular edentulous ridge were fabricated in heat-cured acrylic resin. Heat-cured resin was used for fabricating the models as its modulus of elasticity is closer to natural cancellous bone.[9],[10],[11] The models were designed with a slit in the center of the base to simulate the L-shape and the flexion of the mandible. Implant analogs of dimension 3.75 mm × 11.5 mm were placed in the following tooth positions; 36, 33, 43, 46. UCLA abutments were fixed to the implants. Implant-supported screw-retained fixed prostheses were fabricated in cobalt-chromium alloy and were fixed to the implants with a torque of 35 Ncm using torque wrench. Strain gauges were bonded to the acrylic models mesial and distal to each implant to record peri-implant strain on application of load. Six similar models were fabricated. 400 N of load was applied over the prostheses for a duration of 10 s using universal testing machine, and strain was measured, as in a previous study done by Vasconcellos et al.[6] All the prosthesis were sectioned at the area between 36 and 33, 33 and 43, and 43 and 46 using 35 μm thin carborundum disc, and strain were measured after application of load on the sectioned prosthesis.[12] Specimens were reunited under three groups, namely, arc welding, soldering, and laser welding after which they were subjected to load and strain were measured. The results were subjected to one-way ANOVA to detect statistically significant difference. When sectioning and reuniting of the superstructure was done, a significant difference was observed in the magnitude of strain between the one-piece cast method and various uniting methods. In the present study, lowest mean strain values were observed in models with sectioned prostheses (125–230 microstrains) in all the strain gauges. Whereas, models before sectioning of the prostheses showed the highest mean strain values (435–890 microstrains). Among the three joining techniques, lowest mean strain values were observed when the sectioned prostheses was reunited using laser welding technique (173–260 microstrains) whereas the prostheses reunited by arc welding showed the highest mean strain values (250–385 microstrains). Mean strain values for prostheses reunited by soldering were found to be 227-335 microstrains. Similar results were obtained in a study done by Watanabe et al. in which they compared the peri-implant strain generated by frameworks fabricated by one-piece casting and soldering.[2] Higher mean strain values were obtained in frameworks fabricated by one-piece casting method when compared to frameworks which were sectioned and then reunited using soldering technique. Mendes et al. also observed higher strain in one-piece casting (−355 microstrains) when compared to soldering technique (−0.698 microstrains).[13]

Costa et al. conducted a study to compare the misfit of framework fabricated by one-piece casting and cast in sections followed by laser welding and brazing.[14] Based on the results of the study, they concluded that less distortion in framework was observed when they were cast in sections and reunited by laser welding. They stated that the probable reason for the least strain generated by employing laser welding technique to reunite the sectioned prosthesis could be a small heat affected zone in the metal and the lesser amount of material added to the welded region which reduces the volume of metal that is going to contract on cooling, thus leading to less distortion of welded framework. Whereas in soldering and arc welding technique, greater heat affected zone is formed in metal causing more distortion when compared to laser welding technique. Barbi et al. also conducted a study to compare three different joining techniques, namely, laser welding, brazing, and tungsten inert gas welding by measuring the resulting marginal misfit in a simulated prosthetic assembly.[15] He concluded that the method used for joining Co-Cr prosthetic structures had an influence on the resulting passive fit. Frameworks joined by the tungsten inert gas method produced better mean results than did the brazing or laser welding method. The fit of a framework is determined by the impression method and the material, the dimensional stability of master cast and the fabrication process of the prostheses.[16],[17],[18] The latter is especially important when fabricating a framework by means of lost-wax method. Wax has the highest coefficient of thermal expansion of all dental materials and its dimensional stability is subject to any temperature changes.[19] During investing and casting, distortion occurs which are difficult to eliminate. If an appropriate protocol is followed, the distortion caused by the aforementioned factors is probably small and clinically insignificant.[20],[21] However, a combination of distortion in different dimensions can result in significant misfit at the abutment-implant interface which can generate strain in the bone around the implants.[22],[23] Barbosa et al. stated that any misfit of the prosthesis in relation to the implant will generate external stresses in the prosthesis, implant, and bone, and a rigid and accurate connection between prosthesis and implant is needed for the success of implant-supported prosthesis.[24] The results of the present study are consistent with the concept that it is unlikely that a perfect passive prosthesis might exist because the act of torque application transfers some strain to the abutment and/or bone.

According to the methodology used and based on the results obtained, it was concluded that highest strain value was observed in all strain gauges when single-unit prosthesis was subjected to load whereas least strain was observed when the prosthesis was sectioned and then subjected to load. Increase in strain value was observed in the strain gauges when the sectioned prosthesis was joined by any of the methods and subjected to load, but the values were below those obtained with single-unit prosthesis. Based on the results of this study, the hypothesis of the study was rejected and it was recommended that any long-span implant prosthesis should be sectioned and then reunited preferably by laser welding technique to control the peri-implant strain generated in the surrounding bone.


  Conclusions Top


The following conclusions were drawn from the present study:

  1. Highest strain value was observed in all the strain gauges when single-unit prostheses were subjected to load whereas least strain was observed when the prostheses were sectioned and then subjected to load
  2. Increase in strain value was observed in the strain gauges when the sectioned prostheses were joined and subjected to load, irrespective of the technique used (arc welding, soldering, and laser welding), but the values were below those obtained with single-unit prostheses
  3. Among the three techniques used for the joining of sectioned prostheses, least strain was observed in all the strain gauges when laser welding was used whereas highest strain was observed when arc welding technique was used
  4. Long-span implant prosthesis has to be sectioned and united to control the strain generated in bone around the implants.


Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.

 
  References Top

1.
Tashkandi EA, Lang BR, Edge MJ. Analysis of strain at selected bone sites of a cantilevered implant-supported prosthesis. J Prosthet Dent 1996;76:158-64.  Back to cited text no. 1
[PUBMED]    
2.
Watanabe F, Uno I, Hata Y, Neuendorff G, Kirsch A. Analysis of stress distribution in a screw-retained implant prosthesis. Int J Oral Maxillofac Implants 2000;15:209-18.  Back to cited text no. 2
[PUBMED]    
3.
Sahin S, Cehreli MC. The significance of passive framework fit in implant prosthodontics: Current status. Implant Dent 2001;10:85-92.  Back to cited text no. 3
[PUBMED]    
4.
Kunavisarut C, Lang LA, Stoner BR, Felton DA. Finite element analysis on dental implant-supported prostheses without passive fit. J Prosthodont 2002;11:30-40.  Back to cited text no. 4
[PUBMED]    
5.
Luk HW, Pow EH, McMillan AS, Hui CF. A double casting technique to minimize distortion when constructing fixed partial dentures on implants. J Prosthet Dent 2004;91:93-6.  Back to cited text no. 5
[PUBMED]    
6.
Vasconcellos LG, Nishioka RS, Vasconcellos LM, Nishioka LN. Effects of axial loads on implant supported partial fixed prosthesis by strain gauge analysis. J Appl Oral Sci 2011;19:610-5.  Back to cited text no. 6
    
7.
Himmlová L, Dostálová T, Kácovský A, Konvicková S. Influence of implant length and diameter on stress distribution: A finite element analysis. J Prosthet Dent 2004;91:20-5.  Back to cited text no. 7
    
8.
Eskitascioglu G, Usumez A, Sevimay M, Soykan E, Unsal E. The influence of occlusal loading location on stresses transferred to implant-supported prostheses and supporting bone: A three-dimensional finite element study. J Prosthet Dent 2004;91:144-50.  Back to cited text no. 8
    
9.
Natali AN, Pavan PG, Ruggero AL. Evaluation of stress induced in peri- implant bone tissue by misfit in multi- implant prosthesis. J Dent 2006;22:388-95.  Back to cited text no. 9
    
10.
Rani I, Shetty J, Reddy V. A comparison of peri-implant strain generated by different types of implant supported prostheses. J Indian Prosthodont Soc 2017;17:142-8.  Back to cited text no. 10
[PUBMED]  [Full text]  
11.
Vidya Bhat S, Premkumar P, Kamalakanth Shenoy K. Stress distribution around single short dental implants: A Finite element study. J Indian Prosthodont Soc 2014;14:161-7.  Back to cited text no. 11
    
12.
Tiossi R, Falcão-Filho H, Aguiar FA Jr., Rodrigues RC, Mattos Mda G, Ribeiro RF, et al. Modified section method for laser-welding of ill-fitting cp Ti and Ni-Cr alloy one-piece cast implant-supported frameworks. J Oral Rehabil 2010;37:359-63.  Back to cited text no. 12
    
13.
Mendes SD, Rezende CE, Neto RT, Souse EA, Rubo JH. Effect of framework soldering on the deformation of implant abutments after framework seating: A study with strain gauges. Implants Dent 2013;1:1-6.  Back to cited text no. 13
    
14.
Costa EM, Neisser MP, Bottino MA. Multiple-unit implant frames: One-piece casting vs. Laser welding and brazing. J Appl Oral Sci 2004;12:227-31.  Back to cited text no. 14
    
15.
Barbi FC, Camarini ET, Silva RS, Endo EH, Pereira JR. Comparative analysis of different joining techniques to improve the passive fit of cobalt-chromium superstructures. J Prosthet Dent 2012;108:377-85.  Back to cited text no. 15
    
16.
Yokoyama S, Wakabayashi N, Shiota M, Ohyama T. The influence of implant location and length on stress distribution for three-unit implant-supported posterior cantilever fixed partial dentures. J Prosthet Dent 2004;91:234-40.  Back to cited text no. 16
    
17.
Jackson BJ. The use of laser-welded titanium framework technology: A case report for the totally edentulous patient. J Oral Implantol 2005;31:294-300.  Back to cited text no. 17
    
18.
Savadi RC, Agarwal J, Agarwal RS, Rangarajan V. Influence surface topography and loading condition on stress distribution in bone around implant: A Finite element analysis. Indian Prosthodont Soc 2011;11:221-31.  Back to cited text no. 18
    
19.
Abduo J, Lyons K, Bennani V, Waddell N, Swain M. Fit of screw-retained fixed implant frameworks fabricated by different methods: A systematic review. Int J Prosthodont 2011;24:207-20.  Back to cited text no. 19
    
20.
Silva TB, De Arruda Nobilo MA, Pessanha Henriques GE, Mesquita MF, Guimaraes MB. Influence of laser-welding and electroerosion on passive fit of implant-supported prosthesis. Stomatologija 2008;10:96-100.  Back to cited text no. 20
    
21.
Hollweg H, Jacques LB, da Silva Moura M, Bianco VC, Souza EA, Rubo JH, et al. Deformation of implant abutments after framework connection using strain gauges. J Oral Implantol 2012;38:125-32.  Back to cited text no. 21
    
22.
Tiossi R, Falcão-Filho HB, de Aguiar FA Jr., Rodrigues RC, de Mattos Mda G, Ribeiro RF, et al. Prosthetic misfit of implant-supported prosthesis obtained by an alternative section method. J Adv Prosthodont 2012;4:89-92.  Back to cited text no. 22
    
23.
Yannikakis S, Prombonas A. Improving the fit of implant prosthetics: Anin vitro study. Int J Oral Maxillofac Implants 2013;28:126-34.  Back to cited text no. 23
    
24.
Barbosa GA, das Neves FD, de Mattos Mda G, Rodrigues RC, Ribeiro RF. Implant/abutment vertical misfit of one-piece cast frameworks made with different materials. Braz Dent J 2010;21:515-9.  Back to cited text no. 24
    


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4], [Figure 5], [Figure 6], [Figure 7]
 
 
    Tables

  [Table 1], [Table 2], [Table 3], [Table 4], [Table 5], [Table 6], [Table 7], [Table 8], [Table 9], [Table 10], [Table 11]



 

Top
 
 
  Search
 
Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
Access Statistics
Email Alert *
Add to My List *
* Registration required (free)

 
  In this article
Abstract
Introduction
Materials and Me...
Results
Discussion
Conclusions
References
Article Figures
Article Tables

 Article Access Statistics
    Viewed1571    
    Printed30    
    Emailed0    
    PDF Downloaded152    
    Comments [Add]    

Recommend this journal


[TAG2]
[TAG3]
[TAG4]