Evaluation of flexural strength of Zirconia using three different connector designs: An in vitro study


Mohammed Samiuddin Ahmed, Kareti Mahendranadh Reddy, Y Mahadev Shastry, S Venkat Aditya, P Jayakrishna Babu 
 Department of Prosthodontics, Sri Sai College of Dental Surgery, Vikarabad, Telangana, India

Correspondence Address:
Dr. Mohammed Samiuddin Ahmed
10-3-16/5/D/1, Mehdipatnam, Hyderabad - 500 028, Telangana
India

Abstract

Aim: The aim of this study is to evaluate the flexural strength of zirconia using three different connector designs under vertical and oblique loads. Setting and Design: Invitro - analytical study. Materials and Methods: For simulating zirconia fixed partial prosthesis, a specimen with three octagonal cylinders connected with each other was designed. Each face of the octagon was 3.75 mm ± 0.1 mm, and the total width was 9 mm ± 0.1 mm with a standard connector area of 10 mm2 at cross-section. Three different connector designs, i.e., round, oval, and triangular were milled. Universal testing machine was used to test flexural strength with vertical and oblique forces. Statistical Analysis Used: Intergroup comparison of flexural strength was made using Descriptive statistics (1) one-way ANOVA, Bonferroni's post hoc test (2) Kruskal–Wallis test. The confidence interval was set at 95%, P < 0.05 was considered statistically significant for both the tests. Results: The highest flexural strength was observed in the triangle connector with vertical forces and lowest with oblique forces. Conclusions: Triangle connector design proved to be better than round and oval connectors on the application of vertical loads. Round connector design proved to be better than triangle and oval connector on application of oblique loads.



How to cite this article:
Ahmed MS, Reddy KM, Shastry Y M, Aditya S V, Babu P J. Evaluation of flexural strength of Zirconia using three different connector designs: An in vitro study.J Indian Prosthodont Soc 2020;20:285-289


How to cite this URL:
Ahmed MS, Reddy KM, Shastry Y M, Aditya S V, Babu P J. Evaluation of flexural strength of Zirconia using three different connector designs: An in vitro study. J Indian Prosthodont Soc [serial online] 2020 [cited 2020 Nov 26 ];20:285-289
Available from: https://www.j-ips.org/text.asp?2020/20/3/285/289940


Full Text



 Introduction



The zirconia-based fixed prosthetic restorations demand esthetically pleasing restorations with high strength. For achieving such results technicians design smaller connector size which allows them to give separation of units and naturally appearing embrasures. However, it compromises its overall strength and becomes more prone to fractures.[1] Thus, determining the ideal shape of the connector can be clinically useful.

During mastication, the average force on the posteriors was reported to range between 300 and 880 N. Under horizontal and oblique load, it was found to be 275N. It has been reported that the connector design has an influence on the strength of zirconia prosthetic restorations.[2] Providing a proper cross-sectional dimension and shape of the rigid connectors could become challenging due to the specific, natural shape of the abutment teeth.[3] For Zirconia-based restorations, studies have evaluated the different sizes of the connectors. They recommended that the minimum size of the connector to fabricate a clinically acceptable zirconia restoration is 9 mm 2 at cross-section.[4] The shape of the connector design needs further studies.

 Materials and Methods



The study was approved by Institutional review board Ref No. 616/SSCDS/IRB -E/2017.

Designing of specimens in CAD software

For simulating zirconia 3-unit fixed partial prosthesis, an octagonal specimen with three cylinders connected using different connector configuration were designed at RR Dental Labs Pvt Ltd by Mr Ramana Reddy(CDT), Telangana, India. The octagonal shape facilitates to apply oblique loads at 45°. Each octagonal face was 3.75 mm ± 0.1 mm, and the width of each cylinder was set at 9 mm ± 0.1 mm. The length of the cylinder was 26 mm ± 0.1 mm. Each connector designed had a standard area of 10 mm 2. The separation between each cylinder was 2 mm due to milling limitations. Three different connector designs that were used in clinical scenarios were chosen, i.e., round, oval, and triangular [Figure 1], [Figure 2], [Figure 3] and prepared for milling. Ten samples were tested for vertical loads and 10 for 45° oblique loads for each of the connector designs totaling 60. The total number of samples has been thus divided into the following groups:{Figure 1}{Figure 2}{Figure 3}

Group 1: Round connector for vertical force evaluation-(RV)Group 2: Oval connector for vertical force evaluation–(OV)Group 3: Triangular connector for vertical force evaluation-(TV)Group 4: Round connector for oblique force evaluation-(RO)Group 5: Oval connector for oblique force evaluation-(OO)Group 6: Triangular connector for oblique force evaluation-(TO).

Milling of zirconia specimens

The connector designs were milled out of Zirconia blanks (Shine T) of dimensions 98 mm diameter and 14-mm thickness using 5-axis CAD/CAM milling machine (IMES icore 250i). After milling, the specimens were detached from the mounting frame. The supports were grinded off carefully with a low-speed hand-piece using fine grit diamond bur. All the specimens were sintered according to the manufacturer's instructions (1500°C for 6 h) in a furnace (MIHM-VOGT). The specimens were verified for dimensional accuracy with an electronic caliper to an accuracy limit of 0.1 mm.

Testing on universal testing machine

Specimens were subjected to 3-point bend test using universal testing machine. In triangle shape connector, the base of the triangle was oriented upward for vertical load and turned clock-wise to test 45° oblique load. For oval shape connector, it was oriented such that longer dimension of oval was placed vertically for vertical loads and turned clock-wise to test 45° oblique load. No such orientation was required for testing round connector. The specimens were loaded by means of a mandrel of 6 mm width at a crosshead speed of 1 mm/min placed at the center of the octagonal cylinder.

 Results



Statistical analysis was performed using IBM SPSS version 25.0. Mean flexural strength between groups were analyzed using one-way ANOVA Bonferroni's post hoc test [Table 1] and [Table 2] and Kruskal–Wallis ANOVA with post-hoc analysis using Mann–Whitney tests [Table 3] and [Table 4]. The confidence interval was set at 95%. P [5]

The size, shape, and position of connectors all influence the success of the prosthesis.[6]

Schmitter et al. stated that 9 mm 2 area at cross-section was the ideal connector dimension for zirconia fixed partial prosthesis frameworks. The connector they studied was of 9 mm 2 and was found to be optimum for the strength of the prosthesis and soft tissue around the abutment teeth, which could improve both esthetics and periodontal health.[4] We chose connector area of 10 mm 2 for our study as it depicts an average connector size in the premolar region.

Pantea et al. who compared oval and round shaped zirconia connectors and stated that the behavior of the zirconia-based fixed prosthetic restoration is influenced to a large extent by achieving an optimal connector dimension and crown length. They compared connector of 5 mm 2 and 9 mm 2 and tested for flexural strength and found 9 mm 2 connector of elliptical shape to be significantly stronger. They suggested that the elliptical connector might be stronger due to the wider area of stress distribution.[7]

According to Clausen et al.[8] zirconia prosthesis can be used for posterior restorations. They stated that the ceramic had enough fracture strength to withstand mean masticatory force. In the posteriors, the mean maximum posterior masticatory forces varied from 300 to 880N.

The resultant force exerted due to vertical load on the cuspal inclines of natural teeth were calculated using the formula (N = mg.cos θ). The average cuspal inclination of 37° was taken. In addition, we also calculated resultant force acting on the patient with bruxism habits. For posteriors, it was 276N and for bruxism habituates, it was 963N.

The results of ourin vitro study highlight the fact that the strength of zirconia-based fixed prosthetic restorations is influenced by the proper selection of the rigid connector design for the studied samples and the triangle-shaped connector ensured the best strength. All the connector designs were found capable to withstand vertical and lateral forces exerted during mastication. We suggest that appropriate design should be selected depending on the clinical situation.

From the above, it is evident that the round-shaped connector subjected to oblique forces and the triangular-shaped connector subjected to vertical forces withstood the force better. Round connector withstood oblique forces better due to equal area of force distribution. Triangle connector withstood vertical forces better due to the flat base, which provided better distribution of forces.

An important aspect to be taken into consideration when comparing all these results is the fact that most of the available scientific literature [9],[10],[11],[12],[13],[14],[15],[16],[17],[18],[19],[20] on zirconia strength uses geometric plane samples that do not reflect the actual configuration of a fixed prosthesis, which has curved lines or uneven material thickness, thus leading to an approach different from the ones applicable in clinical situations.

 Conclusions



Within the limitations of the study, we conclude that:

The highest flexural strength was observed in specimens with triangle connectors when force was applied verticallyRound connector design was proved to be better than triangle and oval connector on application on oblique loadsAll the connector designs withstood both vertical and horizontal forces generated during normal mastication.Design of the connector is to be decided by the clinician/technician depending upon the clinical scenario. “One size fits all” cannot be applied in designing the shape and size of the connector.

Limitations

Milling of standardized samples in zirconia is challenging due to the difference in properties of different zirconia blanksMilling calibration changes due to wearing of bur will cause inaccuracies, which might give us false results.

Financial support and sponsorship

RR dental labs.

Conflicts of interest

There are no conflicts of interest.

References

1Tinschert J, Natt G, Mautsch W, Augthun M, Spiekermann H. Fracture resistance of lithium disilicate-, alumina-, and zirconia-based three-unit fixed partial dentures: A laboratory study. Int J Prosthodont 2001;14:231-8.
2Oh WS, Götzen N, Anusavice KJ. Influence of connector design on fracture probability of ceramic fixed-partial dentures. J Dent Res 2002;81:623-7.
3Inan O, Secilmis A, Eraslan O. Effect of pontic framework design on the fracture resistance of implant-supported all-ceramic fixed partial dentures. J Appl Oral Sci 2009;17:533-8.
4Schmitter M, Mussotter K, Rammelsberg P, Stober T, Ohlmann B, Gabbert O. Clinical performance of extended zirconia frameworks for fixed dental prostheses: Two-year results. J Oral Rehabil 2009;36:610-5.
5Gargari M, Gloria F, Cappello A, Ottria L. Strength of zirconia fixed partial dentures: Review of the literature. Oral Implantol (Rome) 2010;3:15-24.
6Mathews MF, Breeding LC, Dixon DL, Aquilino SA. The effect of connector design on cement retention in an implant and natural tooth-supported fixed partial denture. J Prosthet Dent 1991;65:822-7.
7Pantea M, Antoniac I, Trante O, Ciocoiu R, Fischer CA, Traistaru T. Correlations between connector geometry and strength and zirconia-based fixed partial dentures. Materials Chem Physics 2019;222:96-109.
8Clausen JO, Abou Tara M, Kern M. Dynamic fatigue and fracture resistance of non-retentive all-ceramic full-coverage molar restorations. Influence of ceramic material and preparation design. Dent Mater 2010;26:533-8.
9Onodera K, Sato T, Nomoto S, Miho O, Yotsuya M. Effect of connector design on fracture resistance of zirconia all-ceramic fixed partial dentures. Bull Tokyo Dent Coll 2011;52:61-7.
10Kenneth J, Anusavice KJ. Effect of connector design on the fracture resistance of all-ceramic fixed partial dentures. J Prosthet Dent 2002;87:536-42.
11Guazzato M, Proos K, Quach L, Swain MV. Strength, reliability and mode of fracture of bilayered porcelain/zirconia (Y-TZP) dental ceramics. Biomaterials 2004;25:5045-52.
12Studart AR, Filser F, Kocher P, Gauckler LJ.In vitro lifetime of dental ceramics under cyclic loading in water. J Biomaterials 2007;28:2695-705.
13Von Steyern PV, Carlson P, Nilner K. All-ceramic fixed partial dentures designed according to the DC-Zirkon® technique. A 2-year clinical study. J Oral Rehabil 2005;32:180-7.
14White SN, Miklus VG, McLaren EA, Lang LA, Caputo AA. Flexural strength of a layered zirconia and porcelain dental all-ceramic system. J Prosthet Dent 2005;94:125-31.
15Tsumita M, Kokubo Y, Von Steyern PV, Fukushima S. Effect of framework shape on the fracture strength of all-ceramic fixed partial dentures in the molar region. J Prosthodont 2008;17:274-85.
16Larsson C, Holm L, Lövgren N, Kokubo Y, Vult von Steyern P. Fracture strength of four-unit Y-TZP FPD cores designed with varying connector diameter. Anin vitro study. J Oral Rehabil 2007;34:702-9.
17Yilmaz H, Aydin C, Gul BE. Flexural strength and fracture toughness of dental core ceramics. J Prosthet Dent 2007;98:120-8.
18Gurram R, Krishna CH, Reddy KM, Reddy GV, Shastry YM. Evaluating the fracture toughness and flexural strength of pressable dental ceramics: Anin vitro study. J Indian Prosthodont Soc 2014;14:358-62.
19Correia AR, Fernandes JC, Campos JC, Vaz MA, Ramos NV, da Silva JP. Effect of connector design on the stress distribution of a cantilever fixed partial denture. J Indian Prosthodont Soc 2009;9:13.
20Badwaik PV, Pakhan AJ. Non-rigid connectors in fixed prosthodontics: Current concepts with a case report. J Indian Prosthodont Soc 2005;5:99-102.